Tristel ULT is a fast, easy to use, high-level chlorine dioxide disinfectant foam for ultrasound probes. It delivers FDA-cleared HLD in
just 2 minutes without soaking baths or reprocessing equipment. Simply apply the foam, let it sit for the full contact
time, then wipe clean.
Tristel ULT uses Tristel's proprietary chlorine dioxide (ClO₂) chemistry. The dual-compartment dispenser contains Base
Solution (citric acid) and Activator Solution (sodium chlorite); pressing the pump mixes them to generate chlorine
dioxide for use. Designed not to form hazardous by-products or halogenated compounds.
Tristel ULT Key Benefits
- High-level disinfection in 2 minutes at room temperature, for fast, compliant turnaround.
- Proven effectiveness including HPV16/18 and HIV per manufacturer efficacy statements.
- Point-of-care friendly and ready to use, with a compact, portable workflow instead of reprocessing equipment.
- Works for endocavity probes and general transducers when used as directed after cleaning.
- Built-in dispensing system for consistent dosing, designed to help ensure proper application.
When to Use
Use Tristel ULT when you need high-level disinfection for ultrasound probes as part of your standard reprocessing
workflow, especially when fast turnaround matters. It's intended for cleaned, reusable, non-lumened ultrasound probes,
including:
- Endocavity probes such as transvaginal and transrectal probes between patients.
- Skin-surface transducers that may contact non-intact skin during use.
- Point-of-care workflows where you don't want (or don't have) soaking baths or reprocessing equipment. Tristel ULT is
designed to be used close to the exam room.
- Any time your protocol calls for HLD with a 2-minute contact time at room temperature, used exactly per the Directions
for Use (foam applied using Tristel ULT wipes, then residue removed).
Application Instructions
Before you begin: Start with a thoroughly cleaned probe, check the "use by" dates, and wear gloves and safety glasses.
If it's the first time using a new bottle, prime the pump with 4 depressions.
- Pull one dry Tristel ULT Wipe from the tub and close the cap tightly (one wipe per probe/transducer).
- Lay the wipe in your palm and apply 4 doses of Tristel ULT Foam onto the wipe.
- Gently close your hand around the wipe and wait 10 seconds. Don't squeeze it.
- Wipe the probe end-to-end 4 times, covering the entire surface (including shaft and handle) and keeping it visibly wet, especially around ridges/crevices.
- Leave the probe undisturbed for 2 minutes to complete the contact time.
- Remove residue. Put on clean gloves and use a clean Tristel ULT Wipe to thoroughly wipe away all foam residue.
- Store the disinfected probe per your facility protocol and the probe manufacturer's IFU.