Infa-Cap III (34-38 cm head circumference) (Brown) - These caps will fit newborn infants to babies about the age of 9 to 12 months. These caps have an abbreviated electrode placements consisting of F1, F2, C3, C4, 01,02, T3, T4, Cz and ground.
Infa-Cap III can be used for recording premature babies by using E6 Disposable Sponge Disks on all of the electrodes.
Infa-Caps are an EEG electrode application technique. They are made of an elastic spandex-type fabric with recessed, pure tin electrodes attached to the fabric
Sizes in age are approximate; the patient should always be measured to determine proper cap size.
Use only Ivory or Palmolive liquid detergent for washing your cap. Other soaps and detergents can leave a residue on the electrode surface, causing artifacts and/or impedance problems.
Use only ECI Electro-Gel. Other types of paste and gel can cause artifacts and/or high impedance.
Ensure that all accessory electrodes are the same metal as the electrodes in your cap. All standard ELectro-Caps have tin electrodes. Use of accessory electrodes of a different metal can cause a myriad of problems.
Keep all gel, paste, and moisture away from the connector that plugs into your equipment.